DEFINITIONS AND FUNCTIONS
7
77
7.
..
.
ADDITIONAL FUNCTIONS7.1
Stand-by Sequence This is the function not to turn on Alarm Output even if it covers the conditions to acti vate Alarm・・・・
Output at Power ON. This function is valid when the Power turns on at the part of slanting linea
of the following Alarm Output Function Chart.
Alarm Output Hold This function holds the condition of Alarm Output after the Alarm Output has started to activate.・・・・
For release of this function, Power turn on again or setting the position of UNHOLDmake
at Additional Function Setting.
7.2 PVALARM OUTPUT
Alarm Output makes Alarm Output Relay ON/OFF as PV corresponds to Alarm Output PV.
Alarm Output Function Chart・
Deviation High and Low Limit Absolute Value High and Low Limit
!$
<< <<
%& %
△
&
SV 0℃(
I
)
Deviation High Limit Absolute Value High Limit
"D
<<
&
△
&
SV 0℃(
I
)
Deviation Low Limit Absolute Value Low Limit
2E
<<
%%
△
SV 0℃(
I
)
Deviation High a nd Low Limit range Absolute Value High and Low Limit range
3A
<<<<
%& %
△
&
SV 0℃(
I
)
Range of Alarm Output Alarm Output Setting Low Limit Alarm Output Setting Limit:
%
:
&
:High
ensitivity
<
:Alarm Output s
7.3 INPUT DISPLAY RANGE AND SETTING RANGE
(JIS C 1602−1995)CHART 1 DISPLAY RANGE & SETTING RANGE OF THERMOCOUPLE INPUT
Setting Range Display Range Setting Range Display RangeK゚
゚C 0∼1200 −40∼1326 N ゚C 0∼1300 0∼1335
゚F 0∼2200 −40∼2420 ゚F 32∼2350 32∼2435
J ゚C 0∼800 −31∼850 W5Re/ ゚C 0∼2300 0∼2336
゚F 0∼1450 −24∼1563 W26Re ゚F 32∼4200 32∼4236
E ゚C 0∼800 −24∼833 R ゚C 0∼1700 0∼1755
゚F 0∼1450 −16∼1531 ゚F 32∼3100 32∼3192
T ゚C −200∼400 −231∼407 S ゚C 0∼1700 0∼1730
゚F −330∼750 −385∼765 ゚F 32∼3100 32∼3146
(JIS C 1604−1997)CHART 2 DISPLAY RANGE & SETTING RANGE OF R.T.D. INPUT
Setting Range Display Range Setting Range Display Range
()()With Decimal With Decimal
Pt100 ゚C −199∼500 −199∼539 −199.9∼500.0 −199.9∼539.1
゚F −199∼950 −199∼999 −199.9∼950.0 −199.9∼999.9
JPt100 ゚C −199∼500 −199∼529 −199.9∼500.0 −199.9∼529.1
゚F −199∼950 −199∼984 −199.9∼950.0 −199.9∼984.4
CHART 3 DISPLAY RANGE & SETTING RANGE OF CURRENT
Setting Range Display Range
4∼20mADC −1999∼9999/ Displays approx.±12% wider
−199.9∼999.9/ then setting range.
−19.99∼99.99
Commenti su questo manuale